Illumina Miseq
The Miseq is our budget friendly benchtop sequencing system, which is typically used when a relatively small output is needed ex. Amplicon sequencing or whole genome sequencing of small genomes (Bacteria, virus etc.). Several different Sequencing kits are available for the Miseq covering a wide range of outputs, read lengths and prices.
Specifications
Sequencing output per flow cell & runtimes
Miseq Reagent kit v2 | Miseq Reagent kit v3 | |||||
---|---|---|---|---|---|---|
Read length | 1 x 36 bp | 2 x 25 bp | 2 x 150 bp | 2 x 250 bp | 2 x 75 bp | 2 x 300 bp |
Total Time | 4 hours | 5,5 hours | 24 hours | 39 hours | 21 hours | 56 hours |
Output | 540-610 Mb | 750-850 Mb | 4,5-5,1 Gb | 7,5-8,5 Gb | 3,3-3,8 Gb | 13,2-15 Gb |
Mb= Mega bases, Gb= Giga bases
Miseq Reagent Kit v2 Micro | Miseq Reagent kit v2 Nano | ||
---|---|---|---|
Read Length | 2 x 150 bp | 2 x 250 bp | 2 x 150 bp |
Total Time | 19 hours | 28 hours | 17 hours |
Output | 1,2 Gb | 500 Mb | 300 Mb |
Mb= Mega bases, Gb= Giga bases
The outputs listed in the table above, should only be used as a guideline. As the final output of a sequencing run depends on the given sequencing library.
Reads passing QC filter per Flow cell
Miseq Reagent kit v2 | Miseq Reagent kit v3 | Miseq Reagent Kit v2 Micro | Miseq Reagent kit v2 Nano | |
---|---|---|---|---|
Single-end Reads | 12-15 million | 22-25 million | 4 million | 1 million |
Paired-end Reads | 24-30 million | 44-55 million | 8 million | 2 million |
Quality Scores
Miseq Reagent kit v2 | Miseq Reagent kit v3 |
---|---|
> 90% bases higher than Q30 at 1 × 36 bp | > 85% bases higher than Q30 at 2 × 75 bp |
> 90% bases higher than Q30 at 2 × 25 bp | > 70% bases higher than Q30 at 2 × 300 bp |
> 80% bases higher than Q30 at 2 × 150 bp | |
> 75% bases higher than Q30 at 2 × 250 bp |
The Q-Score predicts the probability of an error in the base calling. A Q-score of 30 is equal to a base call accuracy of 99, 9% or in other words an incorrect base call for 1 in a 1000 bases.