Lasers
- Coherent Sapphire 488 nm, air cooled laser (13-20 mW)
- 635 nm JDS Uniphase HeNe, air cooled laser (10-20 mW)
- 407 nm Point Source Violet solid state laser (10-17 mW)
Optical system
- Detection of 14 FL-parameters
- Detection of forward scatter light (FSC) with photo diode
- Direction of side scatter light (SSC) and fluorescence signal towards PMTs
Blue laser detection (488 nm)*
- 502 LP, 530/30 nm FITC
- 556 LP, 585/42 nm PE
- 610 LP, 616/23 nm PE-Texas Red
- 655 LP, 695/40 nm PE-Cy5.5
- 735 LP, 780/60 nm PE-Cy7
Red laser detection (640 nm)*
- 660/20 nm APC
- 690 LP, 730/45 nm Alexa Flour 700
- 735 LP, 780/60 nm APC-Cy7
Violet laser detection (405 nm)*
- 431/28 nm BV421
- 502 LP, 525/50 nm BV480 and BV510
- 595 LP, 610/20 nm BV605
- 630 LP, 660/20 nm BV650
- 690 LP, 720/30 nm BV711
- 750 LP, 780/60 nm BV786
We do have additional bandpass filters that, on request, can be used in the instrument.
Check the reference guide including a guide to setting up the color combination or use the more comprehensive list of available fluorochromes.
* Each color is just an example, you should ensure that you are using the right colors by checking the filter setting.
Sort Collection Devices
-
Two-way sorting into the following sample tube collection devices: microtubes, 12x75 mm, and 15mL
-
Four-way sorting into the following sample tube collection devices: microtubes, 12x75 mm
-
Automatic Cell Deposition Unit (ACDU) option allows for slide and plate sorting into 6, 24, 48, 96, and 384 plates
Sample Input Sizes
-
Microtubes, 12x75 mm, and 15mL
Temperature Control
-
Sample input, software-adjustable: 4, 20, 37 and 42 degrees C
-
Sample output for sort collection: water recirculation unit with variably settings